Mobi
-
Microplate Spectrophotometer
Product Introduction
The Mobi measures light absorption during biochemical reactions, allowing for accurate concentration analysis of samples using microplates.
- Key Features
-
- - Monochromator with selectable wavelengths in 1 nm increments from 200 nm to 999 nm
- - Compatible with microplates ranging from 6 to 384 wells
- - Supports End-point, Kinetic, Spectral Scanning, and Well-Area Scanning measurements
- - Integrated detector and electronics ensure low-noise measurements
- - Equipped with a Xenon flash lamp to extend operating time
- - Options available for 2 μL and 16 μL microvolume plates for nucleic acid quantification
- - Optional incubation and shaking features
- Applications
-
- 01 Nucleic acid quantification
- 02 Protein quantification
- 03 Purity measurements (260/280 and 260/230 ratios)
- 04 ELISA
- 05 Enzyme kinetics
- 06 Cell proliferation
- 07 Microvolume analysis
Product Specifications
| General Specification | |
|---|---|
| Detection Modes | Absorbance |
| Reading Methods | End-point, Kinetic, Spectral Scanning, Well- Area Scanning |
| Microplate Types | 6 to 384 Well Microplate |
| Incubator | Optional (Temperature Control and Shaking) |
| MicroVolume Plate | Optional (2㎕, 16 plate) |
| Power | 100 - 240 V AC, 50/60㎐ |
| Dimensions | 396 x 364 x 207㎜ |
| Weight | 9.0㎏ |
| Reader Specification | |
|---|---|
| Light Source | Xenon Flash Lamp |
| Detector | Photodiode |
| Wavelength Selection | Monochromator |
| Wavelength Range | 200 - 999㎚ (1㎚ Increments) |
| Monochromator Bandwidth | 5㎚ |
| Dynamic Range | 0 - 4.0 OD |
| Resolution | 0.0001 OD |
| Wavelength Accuracy | ±1.0㎚ |
| Wavelength Repeatability | ±0.1㎚ |
| Pathlength Correction | Available |
| OD Accuracy | ±1% at 2.0 OD, 3% at 3.0 OD |
| OD Repeatability | ±1% at 2.0 OD |
| OD Linearfty | ±1% from 0 to 3.0 OD |

_f0c5d.png)

